首页
1
商品介绍
2
LM200 Cryogenic Mill3
177

LM200 Cryogenic Mill

With the rapid development of science and technology, DNA detection and polymer analysis are becoming more and more important. In order to maintain the original compone

Application
It is applied to samples with strong toughness and thermal sensitivity that are difficult to be ground and crushed at room temperature, such as rubber, plastic, food, mineral, animal and plant tissue analysis, drug, biochemical DNA detection, cell disruption and other fields.

Typical sample
polymers, rubber, textile materials, grain particles, hair, nails, skin, bones, muscle tissue, etc.

Working Principle

Size reduction is taken place by electromagnetic oscillation impact. The sample and the metal impactor are loaded into the grinding jar together. The grinding jar is sealed with plug head by the two ends and put into the electromagnetic vibration drum coil. Electromagnetic coil with grinding jar immersed in liquid nitrogen to be pr-cooled. The whole grinding process is carried out in liquid nitrogen (-196 ℃). The magnetic force generated by the coil drives the impactor to move horizontally, thus continuously impact on then grind the sample that has been frozen and embrittled.  

Features and advantages

○Electromagnetic vibration impact guarantees fast grinding speed, high precision, good uniformity and repeatability;
○The liquid nitrogen automatic filling system automatically supplement the liquid nitrogen consumed in the process of sample freezing and grinding to ensure the stability and safety;
○Industrial touch screen control panel, impact frequency, pre-cooling time, impact grinding time, cycle grinding times can be set, easy to operate and reliable;
○Safety design: when the cover of the instrument is open, the instrument cannot be started; Only when the instrument cover is closed and locked can the instrument be started;
○Streamlined and ergonomic design, simple and convenient operation;
○When the liquid nitrogen level is lower than the set value, the liquid sensor will alarm and automatically replenish the liquid nitrogen; and the instrument stops running to protect the electromagnetic coil from damage.

* Information on this site is provided for information purposes only and the contents hereof are subject to change without notice.

Single grinding time 2min
The shortest pre-cooling time 4.5min
Batch capacity 0.1-100g
Pre-cooling chamber 200ml
Grinding chamber 200ml
Grinding jars 4groups(4×3=12pieces)jars of 3ml/4 jars of 20ml /one jar of 100ml/one  jar of 200ml
Liquid nitrogen filling system Manual/ automatic optional
Impact frequency 5-15Hz
Rated power 450W
Power supply 220V,50Hz/60Hz
Instrument size 510* 440*450mm
Package size 720*605*680mm

 